Abstract

Transcatheter aortic valve implantation (TAVI) is performed through a retrograde transfemoral approach in approximately 80-90 % of cases thanks to the improvements in delivery catheter prole, size and steerability compared with the rst generation devices. The aim of this review article is to describe the challenges of transfemoral TAVI and the options to overcome them. The difculties may be related to the access itself or the placement of the valve using the transfemoral route. Comprehensive patient screening using multislice computed tomography and crossover techniques to prevent bleeding should result in low complication rates even for fully percutaneous procedures. Horizontal ascending aorta and severely calcied aortic valves remain a challenge for retrograde valve crossing, device advancement and accurate positioning during deployment. The buddy balloon technique is a simple option in the case of difcult aortic valve crossing with a delivery catheter, whereas an antegrade approach using the transapical route is a reasonable alternative in such anatomy.

Disclosure:Stephane Noble is a proctor for Medtronic CoreValve. Marco Roffi has no conflicts of interest to declare

Received:

Accepted:

Correspondence Details:Stephane Noble, Interventional Cardiology Unit, University Hospital of Geneva, Geneva, Switzerland. E: Stephane.noble@hcuge.ch

Copyright Statement:

The copyright in this work belongs to Radcliffe Medical Media. Only articles clearly marked with the CC BY-NC logo are published with the Creative Commons by Attribution Licence. The CC BY-NC option was not available for Radcliffe journals before 1 January 2019. Articles marked ‘Open Access’ but not marked ‘CC BY-NC’ are made freely accessible at the time of publication but are subject to standard copyright law regarding reproduction and distribution. Permission is required for reuse of this content.

Since the first human transcatheter aortic valve implantation (TAVI)performed through an antegrade transfemoral approach (i.e. femoral venous access, transseptal puncture advancement of the device to the left ventricle and antegradely through the aortic valve), the venous approach has been abandoned and different arterial access routes such as the transfemoral, transapical, subclavian or direct aortic have been developed. At present, TAVI is performed through a retrograde transfemoral approach in approximately 80-90 % of cases thanks to the improvements in delivery catheter prole, size and steerability compared with the first generation devices. Furthermore, transfemoral TAVI has the advantage of offering the option for a fully percutaneous procedure, performed under local anaesthesia with mild sedation, via a puncture of the common femoral artery using the Seldinger technique and a closure device used with a pre-closure technique to achieve haemostasis. The aim of this review article is to describe the challenges of transfemoral TAVI and the options to overcome them. The challenges may involve either the access itself or the valve insertion and implantation.

Challenges in the Vascular Approach of Transfemoral Transcatheter Aortic Valve Implantation
In order to successfully perform transfemoral TAVI, the operators should not only have appropriate knowledge on the devices they use and good training in femoral access and management of its specic complications, but they should also perform comprehensive screening of patients and careful evaluation of vascular access. Genereux et al. showed in a post hoc analysis of the 419 transfemoral patients from the Placement of AoRTic TraNscathetER Valves trial (PARTNER A and B) that vascular complications adversely impacted 30-day and one-year outcomes.1 In this analysis, major vascular complications (the most frequent being vascular dissection, vascular perforation and access site haematoma) occurred in as often as 15.4 % of cases with the use of the first generation Edwards SAPIEN® valves (Edwards Lifesciences Corporation, Irvine, California, US) and were associated with bleeding events, transfusions and increased mortality. Importantly, major vascular complications were associated with more than a fourfold increase in 30-day mortality.1 However, as shown by the Vancouver single-centre experience (137 patients), the incidence of major vascular complications can be dramatically reduced over time. Indeed, major vascular complications (22/24 were iliofemoral) in this centre fell from 8 % in 2009 to 1 % in 2010, thanks to increased operator experience, sheath size reduction (22 and 24 French size [Fr] for the first Edwards SAPIEN generation as opposed to 19 and 18 Fr for the second generation) and improved screening using multislice computed tomography (MSCT).

Furthermore, distinction between major and minor vascular complications is of importance since in the PARTNER trial, vascular subanalysis minor vascular complications, which occurred in 11.9 % of cases, were not associated with higher mortality, major bleeding, transfusions or need for dialysis at 30 days and one year.1

Vascular Screening
Using angiography and MSCT comprehensive vascular screening should determine the degree and localisation of calcications, the presence of relevant iliofemoral atherosclerotic disease, major tortuosity and the minimal lumen vessel diameter that must be at least >6 millimetres (mm) to accommodate an 18 Fr sheath. MSCT is very important in assessing the distribution of vascular calcication. Indeed, in the presence of circumferential calcication and/or tortuosity, the minimum iliofemoral arterial diameter should be larger (preferably >6.5 mm). In the setting of tortuous iliofemoral vessels the feasibility of transfemoral approach can be determined by the ability to straighten the tortuous vessel with a stiff wire (see Figure 1). Vavuranakis et al. developed an index to dene the signicance of vessel tortuosity - the Total arterial Tortuosity/Access vessel Diameter ratio (TT/AD). This ratio corresponds to the sum of angles from the puncture site divided by the minimum femoral artery diameter at the puncture site. In their analysis, a TT/AD >27.7 predicts major vascular complications with 50.8 % sensitivity and 70.6% specicity. This index, although not commonly used in clinical practice, supports the concept that larger arteries are less prone to vascular complications.2

Different 18 Fr Sheaths Used for the Medtronic CoreValve
Different 18 Fr sheaths are available when implanting a CoreValve® (Medtronic Inc, Minneapolis, Minnesota, US). The Cook sheath (Cook Medical, Indiana, US) is the most commonly used. It is robust enough to allow recapture of a partially deployed valve. The latter manoeuvre is more challenging when using the St Jude 18 Fr sheath (ST Jude Medical, Minnesota, US), which is softer and with a better prole, but has the tendency to invaginate during recapture of a partially deployed CoreValve.

The recently available SoloPath™ Balloon Expandable sheath (Terumo Corporation, Japan) represents an interesting option in the presence of difcult vascular anatomy. The concept of this sheath is a reduced prole when folded (outer diameter of 13 Fr) allowing facilitated vessel entry, it is subsequently expanded by balloon ination up to an inner and outer diameter of 19 Fr and 22 Fr, respectively. It provides enhanced manoeuvrability, utilises the natural elasticity of vessels and thus enables delivery through difcult anatomy.3,4 Also, it is designed to collapse upon withdrawal. The SoloPath sheath is an interesting option for borderline iliofemoral access in terms of vascular diameter. However, caution should be paid in the presence of tortuous vessels because of a possibly limited kinking resistance observed in our preliminary experience.

Lubrification with Propofol
To facilitate large bore sheath insertion in calcied vascular access, lubrication of the sheath with sterile propofol may be of help. This tip and trick was also successfully described in the case of stent delivery failure in coronary arteries.5 Indeed, propofol is a short-acting intravenous hypnotic/amnestic agent with lubricating properties, as it is an emulsion of 1.0 % propofol, 10.0 % soybean oil, 1.2 % puried egg phospholipids, 2.25 % glycerol and sodium hydroxide.

Pre-closure Technique Using the Prostar 10XL Device
Surgical femoral cut-down may at first glance seem safer than a fully percutaneous approach. However, surgery may be associated with adverse events such as haematoma, seroma, lymphoceles and infection, and may lengthen the intervention, the timing to ambulation and hospital stay.6 On the other hand, a fully percutaneous approach has been used for many years in the eld of endovascular aortic aneurysm repair and has the advantage of being performed under local anaesthesia with mild sedation. At present, with the reduction of sheath size, many centres favour this latter approach.

A)Iliofemoral Tortuosity Inside the Vessel B)Straightened by Hi-Torque SupraCore Wire
Crossover Approach. A) Endovascular Snare System B) Regular Wire C) Mammary Catheter
When using a percutaneous approach, a closure device - either the Prostar 10XL device (Abbott Vascular, Santa Clara, California, US) or three 6 Fr Perclose devices (Abbott Vascular)7 - should be inserted before the large sheath; the so-called pre-closure technique. Of note, both devices are only approved for closure of 10 Fr sheath holes by the Food and Drug Administration in the US. However, 24.9 % of the PARTNER A and B transfermoral procedures were performed with a fully percutaneous approach using the Prostar XL device. Furthermore, in 2009, Prostar 10XL received a CE Mark and Canada Health approvals to close puncture sites when sheaths up to 24 Fr are used.

At the beginning of the procedure, the sutures of the Prostar device are deployed at the arteriotomy site without tying the knots. At the end of the procedure, the sutures of the pre-closing system are knotted while retrieving the large sheath using either the crossover balloon technique (see below) or with a crossover access to the external iliac artery ready in case of incomplete haemostasis. Failure of the Prostar XL device seems to be associated with obesity, small-diameter vessels with signicant calcication or a high puncture site.8-10 The failure rate of the Prostar device has been reported to be between 7.4 and 15.0 % in TAVI and endovascular aortic aneurysm repair series.6,10-12 In a series of 142 patients from Paris, the rate of failure clearly decreased from early experience (14.3 %) to later experience (4.3 %) whereas the total failure rate was 9.4%.10 Interestingly, in this series, all Prostar failures could be managed without surgical intervention, namely with prolonged balloon ination (3.6 %) or covered stent implantation (3.6 %) and no Prostar failure led to a major vascular complication according to the Valve Academic Research Consortium (VARC) criteria.10

Crossover Balloon Occlusion Technique
Some operators routinely advance a peripheral balloon in the external or common iliac artery from the contralateral access and inate it when retrieving the sheath and knotting the sutures - the so-called crossover balloon occlusion technique. This approach permits bleeding control and offers a bloodless eld for percutaneous or surgical repair if needed.9,13 In our experience, the need to inate a balloon to achieve haemostasis is rarely needed and we therefore do not use it prophylactically. However, we strongly advise having a diagnostic catheter in place in the external iliac artery using a contralateral approach. This can be easily done by advancing a 6 Fr internal mammary artery catheter over a 0.035ÔÇØ wire (usually a stiff Terumo wire, Terumo Inc, Japan). If the aortic bifurcation is narrow-angled and it is not possible to easily advance a wire into the contralateral limb, we advise snaring it (e.g. with EN Snare┬« Endovascular Snare System [Merit Medical Systems, South Jordan, Utah, US]) from the large bore introducer site. This technique may be the only way to 'secureÔÇÖ the contralateral limb in challenging cases (see Figure 2). From the 6 Fr mammary, a nal angiogram is performed to document complete haemostasis at the Prostar site (see Figure 3).

Final Angiography by Crossover Approach. Arrow Showing the Puncture Site
Crossover Final Angiography
It is essential to perform a nal angiography of the vascular access by a crossover approach to verify the haemostasis and to exclude local complications such as dissection or vessel occlusion. Small leaks are not seen from the skin and they may contribute to the formation of a false aneurysm or become the source of delayed major bleeding. If small vascular leakages are seen, prolonged manual compression is usually effective in sealing the leak and a nal angiographic check is required to conrm the sealing. Large leaks or tears often require covered stenting or surgical intervention. In this setting, proximal balloon occlusion is a valuable bridge to endovascular or surgical repair.

In summary, careful patient screening associated with device size reduction and expertise in advanced percutaneous techniques should result in very low complication rates when using the transfemoral approach, even for fully percutaneous procedures. Most local complications can be treated by endovascular methods, surgery is rarely needed.

Challenges in Valve Crossing and Positioning During Transfemoral Transcatheter Aortic Valve Implantation
Despite improvements in delivery catheter prole, size and steerability compared with the first generation devices, retrograde aortic valve crossing during transfemoral TAVI may be challenging especially in the presence of a severely calcied native aortic valve or horizontal ascending aorta.

Sheiban et al. first reported the use of the buddy balloon technique to successfully deliver a transcatheter aortic valve through a severely calcied aortic valve. Despite repeated balloon valvuloplasty, several attempts to deliver a 23 mm Edwards SAPIEN device were unsuccessful, probably due to calcied spiculae and/or recoil. Finally the buddy balloon technique using a 23 x 40 mm NuCLEUS valvuloplasty balloon (NuMed, New York, US) inated at low pressure was successful. The valvuloplasty balloon was used as a shoehorn to enable successful delivery of the Edwards SAPIEN valve.14 Balkin et al. also reported two cases of failure to cross highly calcied aortic valves with the Retroex catheter, used to deliver the Edwards SAPIEN device.15 The buddy balloon technique using an 8 Fr 20 x 40 mm valvuloplasty balloon (OSYPKA AG, Rheinfelden, Germany) allowed successful valve crossing.

At the congress of the European Association of Percutaneous Cardiovascular Interventions (EuroPCR) 2013, we presented a case of retrograde aortic valve crossing with the AccuTrak CoreValve device (Medtronic Inc, Minneapolis, Minnesota, US) requiring the buddy balloon technique to be successful.16 The difference with the previous reports is that we used a 6 Fr compatible 8 x 80 mm Admiral peripheral balloon (Medtronic Inc) instead of a valvuloplasty balloon. The difculty of crossing the native aortic valve in this case was probably due to the horizontal orientation of the ascending aorta. The main advantage of using a 6 Fr compatible balloon is that we do not need to oversize our contralateral access since we routinely insert a 6 Fr introducer in the contralateral femoral side in order to be ready to perform a crossover balloon ination technique in case of major vascular complication when the large sheath is retrieved.

The buddy balloon technique is a simple and easy tip and trick for challenging retrograde valve crossing. Of note, the buddy wire technique was not effective in the report by Sheiban et al.,14 similarly as in our own experience. Finally, the buddy balloon technique may avoid excessive manipulations with the delivery system and wire, and potentially reduce the risk of left ventricle perforation by the wire and embolisation of calcic particles when struggling to cross the valve. Another option in such cases is to convert to transapical approach or even surgical aortic valve replacement. The former is the only TAVI approach using an antegrade crossing of the valve with the device, which is technically easier than the retrograde crossing required with all the other TAVI approaches.

Conclusions
At present the transfemoral approach is the most commonly used approach for TAVI worldwide. Alternative access routes such as direct aortic, subclavian or transapical approaches are of interest when the iliofemoral arteries are diseased, highly calcied or extremely tortuous. Comprehensive patient screening usingmultislice computed tomography and crossover techniques may reduce complication rates when using transfemoral approach, even for fully percutaneous procedures. Horizontal ascending aorta remains a challenge for retrograde valve crossing and device advancement as well as for accurate positioning during deployment. The buddy balloon technique is a simple option in the case of difcult aortic valve crossing with a delivery catheter, whereas the antegrade approach using the transapical route is a reasonable alternative in such anatomy.

References

  1. G├®n├®reux P, Webb JG, Svensson LG, et al., Vascular complications after transcatheter aortic valve replacement: insights from the PARTNER (Placement of AoRTic TraNscathetER Valve) trial, J Am Coll Cardiol, 2012; 60(12):1043-52.
    Crossref | PubMed
  2. Vavuranakis M, Kariori M, Voudris V, et al., Predictive Factors of Vascular Complications after Transcatheter Aortic Valve Implantation in Patients Treated with a Default Percutaneous Strategy, Cardiovasc Ther, 2013 [Epub ahead of print].
    Crossref | PubMed
  3. Dimitriadis Z, Scholtz W, Faber L, et al., Balloon expandable sheath for transfemoral aortic valve implantation: a viable option for patients with challenging access, J Interv Cardiol, 2013;26(1):84-9.
    Crossref | PubMed
  4. Eggebrecht H, Kahlert P, Thielmann M, et al., Usefulness of a novel balloon-expandable vascular sheath for facilitated large-bore arterial access for transcatheter aortic valve implantation, EuroIntervention, 2011;6(7):893-4.
    Crossref | PubMed
  5. Burri L, Toni M, Cook S, Propofol-dip for tricky stent delivery, Cardiovsacular Medicine, 2013;16(5):153-4.
  6. Etezadi V, Katzen BT, Naiem A, et al., Percutaneous suture-mediated closure versus surgical arteriotomy in endovascular aortic aneurysm repair, J Vasc Interv Radiol, 2011;22(2):142-7.
    Crossref | PubMed
  7. Kahlert P, Eggebrecht H, Erbel R, Sack S, A modied ÔÇ£preclosureÔÇØ technique after percutaneous aortic valve replacement, Catheter Cardiovasc Interv, 2008;72(6):877-84.
    Crossref | PubMed
  8. Cockburn J, de Belder A, Brooks M, et al., Large calibre arterial access device closure for percutaneous aortic valve interventions: use of the Prostar system in 118 cases, Catheter Cardiovasc Interv, 2012;79(1):143-9.
    Crossref | PubMed
  9. Genereux P, Kodali S, Leon MB, et al., Clinical outcomes using a new crossover balloon occlusion technique for percutaneous closure after transfemoral aortic valve implantation, JACC Cardiovasc Interv, 2011;4(8):861-7.
    Crossref | PubMed
  10. Hayashida K, Lefèvre T, Chevalier B, et al., True percutaneous approach for transfemoral aortic valve implantation using the Prostar XL device: impact of learning curve on vascular complications, JACC Cardiovasc Interv, 2012;5(2):207-14.
    Crossref | PubMed
  11. Van Mieghem NM, Nuis RJ, Piazza N, et al., Vascular complications with transcatheter aortic valve implantation using the 18 Fr Medtronic CoreValve System: the Rotterdam experience, EuroIntervention, 2010;5(6):673-9.
    Crossref | PubMed
  12. Thomas C, Steger V, Heller S, et al., Safety and Efficacy of the Prostar XL Vascular Closing Device for Percutaneous Closure of Large Arterial Access Sites, Radiol Res Pract, 2013;2013:875484.
    Crossref | PubMed
  13. Sharp AS, Michev I, Maisano F, et al., A new technique for vascular access management in transcatheter aortic valve implantation, Catheter Cardiovasc Interv, 2010;75(5):784-93.
    Crossref | PubMed
  14. Sheiban I, Infantino V, Bollati M, Buddy balloon to deliver a percutaneous aortic valve device: a percutaneous shoehorn?, Catheter Cardiovasc Interv, 2009;74(5):805-7.
    Crossref | PubMed
  15. Balkin J, Silberman S, Almagor Y, Buddy balloon for TAVI, Catheter Cardiovasc Interv, 2010 [Epub ahead of print].
    Crossref | PubMed
  16. Noble S, Rof M, Challenging transfemoral self-expandable bioprosthesis insertion using the buddy balloon technique. Oral presentation in the session: Overcoming TAVI challenges, Presented at: EuroPCR 2013, Paris, France, 21 May 2013.